Citation

PIXL Pinning Pressure Report

This report covers the measurement and evaluation of the pinning pressure achieved by the PIXL Precision Colony Picker, produced by Singer Instruments.
The pinning pressure was measured using a bespoke test procedure, which replicated the standard pinning functionality of the PIXL. Over the span of 80 separate pinning events the PIXL achieved an average pressure of 6.97 ± 0.74 g/mm2, with a maximum recorded pressure of 8.531 g/mm2.